Select your country or region:
SZ7013-PtP
Ni Plated Platinum Sampler Cone with Copper Base
Nickel-plated platinum sampler cones with a copper base combine the superior chemical resistance of a platinum tip, the efficient heat transfer of a copper core, and a protective nickel coating that shields the base from corrosion. While bare copper bases offer excellent thermal conductivity to prevent overheating, they can corrode when exposed to aggressive acid matrices such as hydrochloric acid (>0.5% HCl), aqua regia, or total acid concentrations exceeding 5%. The nickel plating seals the copper base against acid attack and sample deposition without sacrificing heat dissipation—extending cone lifetime, reducing cleaning frequency, and maintaining stable analytical performance. With proper maintenance, the platinum tip can usually be refurbished two to three times to maximize its working life, and once it can no longer be refurbished, its platinum reclaim value can be credited toward any Glass Expansion sample introduction consumables.
Instrument Compatibility
Shimadzu: ICPMS-2040, ICPMS-2050
Specifications
| Type | Sampler |
| Material | Platinum with Nickel plated Copper base |
| Why choose | Top durability, chemical resistance for demanding work and Nickel plated means longer cleaning intervals and suitable with corrosive or high-acid matrices. |
| Good for | For corrosive samples. |
| Cost vs. Life | Highest upfront cost but longest life; can often be refurbished 2–3 times and recycled. |
Consumables for SZ7013-PtP